272
FFT Analysis Modes
13.9 FFT Analysis Modes
13.9.1
Analysis Modes and Display Examples
For the functions of each analysis mode, refer to “13.9.2 Analysis Mode Functions” (p. 289).
Storage
Displays the time-domain waveform of the input signal.
When the window function setting is other than rectangular, the window function is applied to the
time-domain waveform before it is displayed.
Axis
Display type
Description
X axis
Linear
Time-domain display
Displays the value of the time-domain waveform corresponding to the set
frequency range.
Refer to “Relationship Between Frequency Range, Resolution and Number of
Y axis
Lin-Mag
Displays the module waveform
Waveform Example
Window: Rectangular
X axis: Linear
Y axis: Lin-Mag
Histogram
Acquires the amplitude distribution of the input signal.
Main uses:
• To inspect deviations in the amplitude range of a waveform
•
Uses analysis point distribution to ascertain whether a waveform is artificial or natural
About the Functions, refer to “13.9.2 Analysis Mode Functions” (p. 289).
Axis
Display type
Description
X axis
Linear
Displays input level of the input signal.
Y axis
Lin-Mag
Displays analysis data distribution.
Waveform Example
Normal display
X axis: Linear
Y axis: Lin-Mag
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......