Detailed Settings for Modules
8.10.3 Setting Model 8967 Temp Unit
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Mode
Choose the option that matches the type of thermocouple being used.
Selections
Measurement range
Selections
Measurement range
Temp-K
−
200°C to 1350°C
Temp-R
0°C to 1700°C
Temp-J
−
200°C to 1100°C
Temp-S
0°C to 1700°C
Temp-E
−
200°C to 800°C
Temp-B
400°C to 1800°C
Temp-T
−
200°C to 400°C
Temp-W
0°C to 2000°C
Temp-N
−
200°C to 1300°C
RJC (Reference
Junction
Compensation)
When connecting a thermocouple directly to the module, select
[Int]
.
Reference junction compensation is performed within the module.
When connecting through a reference junction device (e.g., a 0°C control tank), select
[Ext]
.
INT
Reference junction compensation is provided within the module.
(Default setting)
(Measurement Accuracy: The sum of the accuracies of the temperature
measurement and the reference junction compensation.)
EXT
Reference junction compensation is not provided within the module.
(Measurement Accuracy: The accuracy of temperature measurement only)
Burn Out
A broken thermocouple wire can be detected during temperature measurement.
Normally when a thermocouple wire breaks, measured values exhibit random instability.
Off
Broken wires not detected.
On
Broken wires detected.
Wire breakage is detected by sensing a miniscule current flow (about 100 nA)
through the thermocouple. If the thermocouple wires are long or composed of a
high-resistance material, set
[Burn Out]
to
[Off]
to avoid measurement errors.
Renew Data
(Data Refresh)
The data refresh rate can be set to Fast, Normal, or Slow.
The default setting is
[Normal]
. This allows stable measurement while eliminating noise.
For quicker response, select
[Fast]
, but note that this will make the measurement
more susceptible to noise. For further improved measurement stability, select
[Slow]
.
Fast
Data is updated approximately every 1.2 ms.
Normal
Data is updated approximately every 100 ms.
(Default setting)
Slow
Data is updated approximately every 500 ms.
8
Utility Functions
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......