220
Save Numerical Calculation Results
4
Enter a filename (if you want to use a different name).
Move the cursor to
[Save Name]
.
Enter a file name. (default setting: MEAS)
Refer to “8.1 Add Comments” (p. 140).
When
[LAN]
is set as the save destination,
[Save
Name]
is ignored and the files will be named in the
previously-determined format.
Refer to “Save Operations (When Setting the Save
Destination to [LAN])” (p. 87).
Save Name
•The maximum number of characters for
[Save
Name]
is 123. The maximum path length
including file name is 255 characters.
•A sequential number starting from 0001 is
automatically appended to the Save Name
(When
[New File]
is selected).
5
Confirm the measurement configuration and
numerical calculation result settings, then start
measurement (Press the
START
key).
After the data is acquired and the numerical
calculation process completes, the numerical
calculation results (text) are automatically saved to
the specified storage media.
Example of Saving Numerical Calculation Results
When you save numerical calculation results, characters or display items used on the instrument are
converted as shown below.
Characters used on the
instrument
Saved characters
2
^2
3
^3
μ
~
u
Ω
~
o
ε
~
e
°
~
c
±
~
+
"Trig Time" ,"No1 Maximum Ch1" ,"No2 Minimum Ch1" ,"No3 Maximum Ch2" ,"No4 Minimum Ch2"
"","V","V","V","V"
"08-04-11 17:40:33.351","+3.00078E-05","+2.12000E-04","+2.00000E-03","+1.30000E-03"
"08-04-11 17:44:25.976","+3.06078E-05","+2.39996E-04","+2.00000E-03","+1.10000E-03"
<When calculation settings are as follows>
Calculation No. 1: Maximum value of analog channel 1
Calculation No. 2: Minimum value of analog channel 1
Calculation No. 3: Maximum value of analog channel 2
Calculation No. 4: Minimum value of analog channel 2
Line 1: Calculation Settings
Line 2: Calculation Result Units
From Line 3: Calculation Results
Results are recorded in the order of the
calculation settings of line 1.
If trigger time is not displayed properly when you open a file with Microsoft Excel
®
, select “Custom” for
[Format] of [Format Cells] and enter “yy-mm-dd hh:mm:ss.000” for Type.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......