87
Save Data
Example 5
Saving data in an existing folder on the media using divided saving
Save to
CF:¥HIOKI_
MR8827¥TEST
Save method
Normal Save
Folder to save Off
(Create a folder named “TEST” on the CF card
beforehand.)
When the storage media is full:
Automatic saving stop.
Example 6
Saving data in an automatically created folder on storage media using divided saving
Save to
CF:¥HIOKI_
MR8827
Save method
Normal Save
Folder to save On
When the storage media is full:
Automatic saving stop.
Save Operations (When Setting the Save Destination to
[LAN]
)
The file is saved according to the method (on the
right figure) specified under
[Save Settings]
using
the 9333 LAN Communicator.
For more information about how to configure save
settings, refer to the instruction manual for the 9333
LAN Communicator.
The destination folders and files are named in the
previously-determined formats and saved as follows:
Specified folder
192.168.1.1
14-01-10
14-01-11
HIOKI 9333
File names will be
“WAVE + hours/
minutes/seconds.”
Waveform files
14-01-20
14-01-21
File names will be
“PRA
number.”
Folder names
will be
“YY-MM-DD.”
File names will be
“WAVE + hours/
minutes/seconds.”
WAVE124012.BMP
Numerical calculation files
PARAMETER.TXT
PARAMETER_1.TXT
WAVE123015.MEM
WAVE123245.MEM
WAVE123245.TXT
Folder name
will be Model
MR8827 IP
address.
TEST
0000AUTO.IDX
0001AUTO.MEM
0001TEST
Up to 5,000 files
0000AUTO.IDX
0000AUTO
0001AUTO
Up to 5,000 folders
0000AUTO.MEM
0000AUTO.MEM
The automatically created folder will be
named “4-digit folder name.”
0000AUTO
(A new folder will be created automatically when 5,000
folders have been saved in the TEST folder.)
HIOKI_MR8827
AUTO014530
0000AUTO.IDX
0001AUTO.MEM
0001AUTO014530
(A new folder will be created automatically when 5,000 folders
have been saved in the AUTO014530 folder.)
Up to 5,000 files
0000AUTO.IDX
0000AUTO
0001AUTO
Up to 5,000 folders
0000AUTO.MEM
0000AUTO.MEM
“AUTO + hours/minutes/seconds”
is created automatically. (Example: 1:45:30)
Folder name created next will be
“4-digit folder name.”
0000AUTO
HIOKI_MR8827
5
Save/Load Data and Manage Files
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......