171
Detailed Settings for Modules
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Level
This is enabled only when
[Mode]
is
[Pulse Width]
or
[Duty].
In pulse width/duty rate measurements, set which level is detected when the
threshold is exceeded.
HIGH
Measures above the threshold value.
(Default setting)
LOW
Measures below the threshold value.
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Smoothing
This is enabled only when
[Mode]
is
[Freq]
or
[RPM].
Set smoothing.
Off
The measured data is recorded as is. (A step waveform.)
(Default setting)
On
The measured data is interpolated so that the waveform is smooth
before it is output. (Upper limit: 10 kHz, further delayed from Off.)
Hold
This is enabled only when
[Mode]
is
[Freq]
or
[RPM].
Set hold for frequency and count.
Off
(1Hz/0.5Hz/
0.2Hz/0.1Hz)
When the frequency reaches the value in the brackets but is not settled,
it is defined as a stopped item and the measurement value becomes 0
Hz (0 rpm).
(Default setting)
On
The value defined last time is retained.
8
Utility Functions
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......