390
Error Messages
Msg
No.
Message
Remedy
References
18
Not enough free space in the
folder to create file.
Either delete files in the saving
destination folder, or change to
another saving destination.
—
19
Folder not empty.
The folder is not empty, be sure
you want to delete.
—
20
The maximum length of a file
name, including its path, is 255
characters.
File names including the path
must not exceed 255 characters.
—
21
Internal error.
An internal error occurred. Check
the storage media.
—
22
No waveform data to save.
Acquire waveform data.
—
24
There is no calculation result.
There is no calculation result.
Print results after performing
calculation.
25
Cannot select this media.
You cannot select this medium
as a destination of auto-saving.
Select another medium.
—
26
Invalid folder path.
Route cannot be specified.
—
27
When executing, select “No” as
the save property.
As other dialogs are displayed,
selective saving cannot be
carried out. Either set selective
saving during execution to
[No]
or close the dialog and re-save.
28
Pressed key invalid (waveform
evaluation).
Unable to perform operation
because waveform evaluation
is enabled. Turn
[Off]
waveform
evaluation.
29
The recording length was
restricted.
—
—
30
Auto-ranging failed.
Check the input signal.
“3.6 Measurement With Automatic
31
AB cursor positions invalid.
The AB cursors overlap. Check
the cursor positions.
32
Zero-adjustment needed.
Perform zero-adjustment.
33
Disabled key.
Close the dialog.
—
34
Invalid key pressed (Overlay)
The key operation is prohibited
because Overlay is enabled. Set
Overlay to
[Off]
.
Over Past Waveforms (Overlay)”
36
No trigger has been set.
Set trigger criteria.
38
Use of logic channels has
reduced analog waveform
resolution from 16 to 12 bits.
Using channels LA - LD reduces
the resolution of CH1 to CH4 to
12 bits. Using channels LE - LH
reduces the resolution of CH17
to CH20 to 12 bits.
39
Auto-balance failed.
Check whether the strain gauge
transducer is unloaded and
connected correctly.
“8.10.4 Setting Model 8969 and
40
Voltage Sag triggering is
disabled. (Valid time base range:
20 μs/div to 50 ms/div)
Voltage Sag triggering can be
used only when the time base is
between 20 μs/div to 50 ms/div
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......