58
Measurement Configuration
Recording Length and Data Samples
Each division of the recording length consists of 100 data samples. The total number of data samples for
a specified recording length is: Set recording length (divisions) × 100 + 1.
However, if the MR8990 Digital Voltmeter Unit is installed, the number of data samples is as follows.
• When only MR8990 is installed
Number of data samples for each division: 50 data samples
Total number of data samples for recording length: Set recording length (divisions) × 50 + 1
• When a mix of MR8990 and other modules are installed
Channels for MR8990 units
Number of data samples for each division: 50 data samples
Total number of data samples for recording length: Set recording length (divisions) × 50 + 1
Channels for other modules
Number of data samples for each division: 100 data samples
Total number of data samples for recording length: Set recording length (divisions) × 100 + 2
Each recording length division consists of 100 pairs of data points, with each pair composed of two
values: the maximum and minimum measured values within each sampling period. The MR8990 data
measured by the recorder function has a resolution of 16 bits.
When recording length is set to
[Cont.]
• The instrument stores the maximum recording length of data back from the point at which measurement
was stopped in its internal memory (80,000 divisions).
• For range settings of 10 ms/div to 200 ms/div, when the printer setting (real-time print) is On, the real-
time print is not performed. Perform printing manually after measurement is completed. (p. 112)
• When Auto-saving is On, the saving is not performed during the measurement. When measurement is
stopped, the data currently stored in the internal memory is saved.
When recording length is changed during measurement
Measurement data already stored in memory is discarded and measurement starts with the
newly set recording length.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......