91
Save Data
7
Advanced settings
Available settings will differ depending on the save type.
Save
types Settings
Description
Setting
-
-
Wave
Binary
Division
(Off, 16 M, 32 M, 64 M)
Select this to divide large files into
several files when saving.
A folder with the specified name
will be created and data saved as
multiple files in that folder.
Wave
Text
Save
Thin
(Off, 1/2 to 1/1000)
Select this to use thinning (sub-
sampling) when saving text data.
Set the thinning amount (the fraction
of the total data to retain).
Screen
Copy
(Hard
copy
of the
display)
Image
Color
(Color, Gray, Mono, Mono [Rev])
Specifies the color palette of the
created image file.
Image
size
(UnCompressed, Compressed)
Determines whether the image file is
compressed.
Save
GUI
(Yes, No)
Determines whether the GUI will be
saved in the image.
Wave
Image
image)
Output
File
(All, 1 to 250)
Specifies the number of files to save.
To specify a range, open the
[Printer]
sheet from the System screen and
set the printing area to
[A-B Wave]
.
(Memory/Recorder Function only)
Wave
Image
Shot
(1 to 60 div)
Specifies the amount of data per file.
If the amount of measurement
data is lower than the entered
number of divisions, only the actual
measurement data amount will be
saved. (Memory/Recorder Function
only)
Calc
Result
Folder
to save
(New File, Existing File)
Determines whether data is saved
under a new file name every time (with
automatically incremented number
if the same file name exists), or
appended to the same file.
8
(For memory function)
When using the memory division function
Select the blocks to be saved
Move the cursor to
[Save Block]
.
All Blks
Saves all blocks with waveforms.
Start-End
Saves all used blocks beginning from the
start block.
This completes the data save settings.
These settings will be used whenever the
SAVE
key
is pressed.
If “Image size” is used when saving a display image file, the image may not be viewable with some
image viewer applications.
5
Save/Load Data and Manage Files
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......