154
Convert Input Values (Scaling Function)
To Select a Clamp Sensor
1. Move the cursor to
[Clamp]
, and select
[Select]
.
The cursor moves to
[Model].
2. Select
[9000~]
.
The cursor moves to
[Clamp].
3. Select
[9018-50]
from the clamp list with
F
key, and select
[Confirm]
.
Units, scaling method, and ratio will be set automatically.
4. Select the same range of the clamp sensor when using the range selection type. Select
[10A]
here.
However, you may need to switch the vertical axis (voltage axis) range to suit actual input values.
For example, to display 0.2 V at full scale, set the vertical display to 20 mV per division (the
instrument's 20 mV/div range)
[A]
Before Scaling
After Scaling
With scaling, signals from the sensor are acquired
as current rather than voltage.
AB cursors and gauges are displayed and printed
as current (Ampere) values.
Refer to “Gauge” (p. 116) or “Cursor AB Value”
When using Model 8969 or U8969 Strain Unit
Example 2
To display data measured with the strain gauge transducer with a rated capacity
of 20 G and a rated output of 1000 µV/V in terms of gees (G).
For the rated capacity and rated output, consult the inspection record of the strain gauge
transducer to be used. Set the rated capacity and rated output so that the following expression
holds:
Setting
Selection
Set
NUM
Unit
G
Rated Cap (Rated
capacity)
20.000
Rated Output
1.0000k
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......