Module Specifications
Operating
environment
As per Memory HiCorder in which Model 8970 is installed
Storage temperature
and humidity
As per Memory HiCorder in which Model 8970 is installed
Standards
Safety EN61010
EMC
EN61326 Class A
Dimensions
Approx. 106W × 19.8H × 196.5D mm (4.17″W × 0.78″H × 7.74″D)
Mass
Approx. 250 g (8.8 oz.)
18.6.6
Model 8971 Current Unit
Accuracy figures assume installation in a Memory HiCorder and operation after a 30-minute warm-up period at 23°C
±5°C (73°F ±9°F) and 80% RH (no condensation).
Product warranty
period
3 years
Guaranteed accuracy
period
1 year
Number of input
channels
2 channels
Applicable current
sensors
Model 9272-10, 9277, 9278, 9279, 9709, CT6841, CT6843, CT6844, CT6845, CT6846,
CT6862, CT6863, CT6865
Hioki current sensors with RM515EPA-10PC (Hirose) installed
(Supported conversion ratios: 20 A/2 V, 50 A/2 V, 200 A/2 V, 500 A/2 V, 1000 A/2 V*
2
)
(Model 9318 Conversion Cable is required to connect a sensor with Model 8971.)
Measurement range
When Model 9272-10 (20 A), 9277, or CT6841 is used:
100 m, 200 m, 500 m, 1, 2, 5 A/div
When Model CT6862 is used:
200 m, 500 m, 1, 2, 5, 10 A/div
When Model 9272-10 (200 A), 9278, CT6843, or CT6863 is used:
1, 2, 5, 10, 20, 50 A/div
When Model 9279, 9709, CT6844, CT6845, CT6846*
2
, or CT6865*
2
is used:
2, 5, 10, 20, 50, 100 A/div
Measurement
accuracy*
±0.65% f.s. (Filter 5 Hz On)
±0.85% f.s. (Filter 5 Hz On, for using with Model 9278 or 9279)
RMS accuracy*
±1% f.s. (DC, 30 Hz to 1 kHz),
±3% f.s. (1 kHz to 10 kHz)
(Sine wave input, filter 5 Hz On, crest factor 2)
Response time*
100 ms (rising from 0% f.s. to 90% f.s.)
Temperature
characteristic*
±0.075% f.s./°C
Frequency
characteristic*
DC Coupling: DC to 100 kHz ±3 dB
AC Coupling: 7 Hz to 100 kHz ±3 dB (low cut-off frequency: 7 Hz±50%)
Noise*
10 mA p-p max. (Highest sensitivity range, with input shorted) (for 20 A/2V range)
Low-pass filter
OFF, 5, 50, 500, 5 k, 50 k±50% (Hz) −3 dB
Input type
Unbalanced input (Not insulated)
Input coupling
AC/DC/GND
Input resistance
1 M
Ω
±1%
A/D resolution
12 bits
Maximum sampling
rate
1 MS/s
Specifications
18
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......