A
18
FFT Definitions
Imaging
When the instrument is set to a measurement frequency range that requires a higher sampling rate than
the maximum capability of the module, intermediate data points are interpolated between successive
data samples. In this case, the time-domain waveform exhibits a stair-step shape. When FFT analysis is
performed in this situation, non-existent high frequency spectral components appear. This phenomena is
called zero-order hold characteristic
imaging
.
The following figures show the time-domain waveform and spectrum of a sine wave applied to the Model
8968 High Resolution Unit.
Spectral Imaging
Time-domain waveform in the 8
MHz frequency range (sampling
frequency = 20 MHz)
The highest sampling frequency of the
Model 8968 is 1 MHz, so the same input
data value is used for each block of 20
samples, resulting in a stair-step waveform.
Spectrum
When FFT processing is performed on a
stair-step waveform, the resulting spectrum
shows non-existent components.
In this case, the spectral components
above 1 MHz / 2 = 500 kHz are theoretically
meaningless.
Time-domain waveform in the 400
kHz frequency range (sampling
frequency = 1 MHz)
Here, the frequency range matches the
sampling frequency of the Model 8968 so
no interpolation is performed on the time-
domain data.
Spectrum
To avoid imaging phenomena when analyzing waveforms with the FFT function, verify the maximum
sampling frequency of the module before measuring.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......