Trouble Shooting
19.1 Trouble Shooting
If damage is suspected, check this section before contacting your authorized Hioki distributor or reseller.
Before Having the Instrument Repaired
If Power and Operating Keys Malfunction
Symptom
Cause or Item to check
Remedy and reference
The display does not appear
when you turn the power on.
Is the power cord disconnected?
Are connections made correctly?
Verify that the power cord is connected
properly.
Keys do not work.
•Is any key being held down?
Verify key operation.
•Is the key-lock state active (Key-Lock
message displayed)?
Deactivate the key-lock state.
(Hold the
ESC
keys for three seconds)
If the Display or Operations Malfunction
Symptom
Cause or Item to check
Remedy and reference
Blank screen
Is the backlight saver active? Press any
key to check.
Turn the backlight saver off.
No waveform appears when
you press the
START
key.
•Is the “Pre-Trigger wait” message
displayed?
•Is the “Trigger wait” message
displayed?
When pre-triggering is enabled, triggering
is ignored until the pre-trigger portion
of the waveform has been acquired.
Recording starts when a trigger occurs.
No changes occur in the
displayed waveform.
•Is the clamp sensor or connection cable
connected correctly?
•Is the vertical axis (voltage axis) range
set properly?
•
Is the low-pass filter enabled?
Verify that the clamp sensor or
connection cable is connected correctly.
Verify the input channel settings.
While measuring with
the memory function, the
displayed frequency is
much lower than the actual
frequency.
Aliasing may be occurring.
Change the timebase to use a faster
sampling rate. Press the
AUTO
key to try
automatic setting.
“3.3.2 Timebase and Sampling Rate”
The displayed waveform
size does not change when
the input range is changed.
Is the Variable function
[On]
?
Turn the Variable function
[Off]
.
“8.6 Set the Waveform Position (Variable
19
Maintenance and Service
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......