85
Save Data
10
Set the save method for when the storage media
runs out of space.
Move the cursor to
[Save method]
.
Normal
Save
Automatic saving stops when the storage
media is full.
Delete
Save
Old files are deleted and automatic saving
is continues when the storage media is full.
(Waveform files only.)
When
[LAN]
is set as the save destination, this
setting is ignored and the save method becomes
[Normal Save]
.
11
Confirm the measurement configuration then start
measurement.
Press the
START
key.
After the data is acquired, the screen image is saved
automatically to the specified storage media.
•
Maximum number of files per folder
The combined maximum for files and folders created in one folder is 5,000.
Refer to “Auto-save Operations (When Setting the Save Destination to Storage Media)” (p. 86).
•
[Preparing files]
display
Sometimes
[Preparing files
] is displayed after measurement starts. This indicates that the instrument
is checking the status of the destination where the file will be saved. The more the files exist in the
destination folder, the longer the preparation takes.
When printing is attempted simultaneously with saving, the printing may override the saving and
[Preparing files]
may be displayed longer.
To clear the Auto Save dialog box
Press the
SAVE
key to toggle dialog box display.
5
Save/Load Data and Manage Files
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......