337
Connect to External Control Terminals
17.1 Connect to External Control Terminals
The method for connecting to the external control terminals is as follows.
Procedure
Cables to connect
Solid wire
Stranded wire
10 mm
Recommended wire
Solid wire: 0.65 mm dia. (AWG22)
Stranded wire: 0.32 mm
2
(AWG22)
Compatible wire:
Solid wire: 0.32 to 0.65 mm dia. (AWG28 to 22)
Stranded wire: 0.08 to 0.32 mm
2
(AWG28 to 22)
Strand diameter: 0.12 mm or more (per wire)
Stripped length:
9 to 10 mm
Button pressing tool:
Flat-blade screwdriver (shaft diameter: 3 mm, tip
width: 2.6 mm)
Connection procedure
Back side
1
2
3
1
Push in the button on the connector with
a flat-blade screwdriver or other tool.
2
With the button held in, insert the cable
into the cable connection hole.
3
Release the button.
The cable is locked.
Terminal
No.
Operation
1
Input external signal and execute the
following
•Start/end measurement
•Print/save data
•Pen Up/Down Operation (for X-Y
recorder operation)
2
3
4
—
5
Output instrument status as a signal
6
7
—
8
Input external signal and set sampling
rate
9
—
10
Output signal when triggering occurs
11
Input external signal as trigger source
17
External Control
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......