110
Auto-print Settings
4
Set the margins in automatic print
Move the cursor to the
[Feed]
.
Yes
Margin (Default setting)
No
No margins*
* There is a margin of about 2 mm.
[Feed]
are also applicable to normal printing.
5
Check the measurement conditions and start the
measurement.
(Press the
START
key.)
When using the Memory Function
Data is acquired for the recording length then printed
automatically.
When using the Recorder Function
Printing is carried out in real time along with recording.
When using the FFT Function
Data is printed automatically after FFT analysis finishes.
To stop printing before it has finished
Press the
STOP
key. Measurement will also stop.
Printing can be paused and restarted using the F key during Real-Time Printing (Recorder
Function).
Move the cursor to the
[Realtime Print]
item, and select On or Off.
When printing is carried out again after stopping, the range selection will be used.
•
When both auto-print and auto-save are enabled, auto-save is executed first.
However, when the Roll Mode function (default setting:
[Auto]
) is used with the Memory function,
auto-print has priority.
•
If AB cursors are on during manual printing after acquiring waveform data, the selected range will be
printed.
Timebase
(Time/div)
Measurement conditions
Printing
Up to 200 ms/div
Memory Function (Roll Mode: Off)
Data is acquired for the recording length
then printed automatically
Memory Function (Roll Mode: On)
Recorder Function (Recording length: Other
than Continuous [Cont.])
Print simultaneously with some delay from
recording.
Recorder Function (Recording length:
Continuous [Cont.])
Printing disabled
More than 500 ms/
div
Memory Function (Roll Mode: Off)
Data is acquired for the recording length
then printed automatically
Memory Function (Roll Mode: On)
Recorder Function
Print simultaneously with recording.
When performing waveform evaluation, the evaluation area will be printed at the same time.
To disable printing of the evaluation area, set waveform evaluation to
[Off]
If
[Output]
is set to
[LAN]
, realtime printing is disabled. The Auto Print prints automatically after data is
acquired for the recording length regardless of the timebase setting.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...