156
Convert Input Values (Scaling Function)
When the strain gauge transducer’s inspection record provides the calibration
factor
Set
[Method]
on the
[Scaling]
sheet to
[Ratio].
Example 3
To display the data measured with the strain gauge transducer with a calibration
factor of 0.001442 G / 1 10
6 strain* in terms of gees (G).
The value of the calibration factor (0.001442 [G]) is set as the conversion ratio.
(* 10
−6
strain =
με
)
Move the cursor to the appropriate channel, and set as follows.
Setting
Selection
Se
NUM
Unit
G
Ratio
(Conversion
ratio)
0.001442 [G]
(Displays as 1.4420m)
Using a Strain Gauge With a Gauge Rate Other Than 2.0
The Model 8969 or U8969 Strain Unit measures the gauge rate as 2.0.
When using other strain gauges, the gauge rate needs to be set as the
conversion ratio.
For example, if the gauge rate is 2.1, the conversion ratio is 0.952 (≈2/2.1).
Example 4
Measure using a strain gauge (gauge rate: 2.1), and display the measured data in units of G.
The scaling (conversion ratio) needs to be calculated to include both gauge rate and physical
value conversions. In this case, the conversion ratio setting is the product of the conversion ratios
of the gauge rate and measurement unit scaling.
The gauge rate component of the conversion ratio is 0.952, and the physical value component is
0.001442*
Conversion Ratio = 0.952 × 0.001442 = 0.0013728
As in Example 3, enter
[0.0013728]
as the conversion ratio.
To convert measurement values to physical values when using a strain gauge, calculate using
the Young's modulus or Poisson's ratio of the measurement object. The conversion method
depends on the conditions in which the strain gauge is used.
Refer to “Appendix 2.5 Scaling Method for Strain Gauges” (p. A8).
This value cannot be changed.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......