362
Module Specifications
Reference junction
compensation
Selectable internal or external (during thermocouple measurement)
Temperature
characteristic
Add (measurement accuracy × 0.1)/°C to measurement accuracy
Renew Data (Data
Refresh)
Selectable from normal, fast, slow
Fast
Normal
Slow
Data Refresh
Rate
Approx.
1.2 ms
Approx.
100 ms
Approx.
500 ms
Wire break
detection
Selectable from On or Off
Input resistance
5 M
Ω
or more (Both when wire break detection is ON and OFF)
Common mode
rejection ratio
80 dB or more (at 50 Hz/60 Hz, with signal source 100
Ω or less
, Fast data refresh setting)
100 dB or more (at 50 Hz/60 Hz, with signal source 100
Ω or less
, Normal data refresh
setting)
Input type
Unbalanced input (floating)
Maximum rated
voltage to earth
300 V AC, DC (between individual input channels and enclosure as well as between
individual input channels)
Measurement category II (anticipated transient over-voltage: 2500 V)
Operating
temperature and
humidity
As per Memory HiCorder in which Model 8967 is installed
Storage
temperature and
humidity
-20°C to 50°C (-4°F to 122°F), 90% RH or less (no condensation)
Operating
environment
As per Memory HiCorder in which Model 8967 is installed
Dimensions
Approx. 106W × 19.8H × 204.5D mm (4.17″W × 0.78″H × 8.05″D)
Mass
Approx. 240 g (8.5 oz.)
Effect of radiated
radio-frequency
electromagnetic
field
±2% f.s. (max.) at 3 V/m
Effect of conducted
radio-frequency
electromagnetic
field
±2% f.s. (max.) at 3 V
Standards
Safety EN61010
EMC
EN61326 Class A
Accessories
Ferrite clamp-on choke (2 pieces)
Options
(sold separately)
Model 9810 Thermocouples (K)
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......