148
Display New Waveforms Over Past Waveforms (Overlay)
8.3 Display New Waveforms Over Past Waveforms
(Overlay)
This applies to the Memory function only.
Displayed waveforms are retained on-screen and overlaid with new waveforms.
• Use this to compare new waveforms with those recorded immediately before. (When the trigger mode is
[Repeat]
or
[Auto]
• There are two overlay methods: automatic overlay during measurement and manual overlay.
Normal display
Waveforms with the Overlay Function
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Status]
sheet.
Move the cursor to
[Overlay].
Off
Overlay disabled.
(Default setting)
Auto
Each time a waveform is acquired, it is automatically
displayed as an overlay.
When the trigger mode is
[Cont.]
or
[Auto]
, waveforms
will be overlaid from measurement start to stop.
Manual
Overlay waveforms on the screen manually. (See the
next section for the method of overlay.)
This mode cannot be used
simultaneously with Roll Mode.
Refer to “When the overlay function
is enabled ([Auto] or [Manual]).”
Manual Overlay (Any Waveform Can Be Retained On-Screen)
Press the
DISP
key to open the Waveform screen.
Move the cursor to
[Overlay].
Overlay
(
F1
key)
The loaded waveform on the screen.
The overlay is displayed until the waveform is cleared.
Clear
(
F5
key)
Clears the screen of all overlaid waveforms. Cleared
waveforms cannot be displayed again.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......