265
Screen Display Settings
13.5.1 Display Running Spectrums
If
[Format]
is set to
[Running spectrum]
, you can observe changes in frequency over time.
Procedure
Press the
STATUS
key to open the Status screen, and then select the
[Status]
sheet.
1
Select the reference data.
Move the cursor to
[Reference]
, and select
[New Data]
.
2
Select the display format.
Move the cursor to
[Format]
, and select
[
Running
spectrum]
.
•
[Running spectrum]
can only be set when
[Reference]
is
[New Data]
.
• The calculation interval (time interval of the running spectrum waveform and the waveform) is not
regulated.
• Averaging cannot be used.
• Calculations that can be analyzed by running spectrum are limited to the following list. When other
calculations are selected or calculation settings are changed after measurement stops, waveforms
will not be displayed.
(Linear spectrum, RMS spectrum, power spectrum, power spectrum density, LPC analysis, transfer
function, cross power spectrum, 1/1 octave analysis, 1/3 octave analysis)
• When running spectrum is displayed, the auto-saving/auto-printing during measurement cannot be
performed.
13
FFT
Function
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......