176
Detailed Settings for Modules
Offset
For DC output: Sets DC voltage.
For sine wave output: Sets the offset voltage.
The output voltage with guaranteed accuracy is the sum of the amplitude and the
offset, between -10 V and +10 V. If the sum of the amplitude and the offset is set
outside the guaranteed accuracy range, parts of the waveform will be clamped to the
upper limit, approxi14 V and the lower limit, approximately -14 V.
-10 V to +10 V
When Off
Sets the terminal status when the output is off.
Open
Opens the output terminals, separating them from internal circuits.
Short
Short-circuits the output terminals, separating them from internal circuits.
Output
Turns waveform output On/Off.
On
Outputs waveform.
Off
Does not output the waveform.
Control
Sets the waveform output.
RUN
Starts output.
PAUSE
Pauses output. While output is paused, the output at the time [PAUSE]
was pressed will be output.
STOP
Stops output.
Method
Selects the control method for waveform output.
Manual
Restricts control of signal output to Signal Generation screen.
Sync.
Augments manual control with signal output in synchronization with
the start and end of measurement.
START
key: Starts output when measurement starts.
STOP
key: Stops output when measurement stops.
Keys
Augments manual control by allowing signal output to be
manipulated using the instrument’s keys.
START
key: Starts output.
STOP
key: Stops output.
Manual Trigger key: Pauses output.
(For details, see the instruction manual of Models U8793, MR8790, and MR8791.)
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......