391
Error Messages
Msg
No.
Message
Remedy
References
41
Channels that are not
measurable will be assigned to
X-Y.
Channel with specification
exceeding
[Used Ch]
is selected.
Change channel selection.
42
There is not enough data in the
memory.
Measure the correct amount data
required for calculation.
43
Stopped.
—
—
44
Current clamp/sensor was
recognized.
—
—
45
Current clamp/sensor was
disconnected.
Check the connection of the
current clamp/sensor.
—
46
Channels may no longer be able
to be used by frequency units
due to logic channel use.
If LA to LE or LD to LH are used,
the frequency units cannot be
used at channels CH1 to CH4 or
CH17 to CH20 respectively.
“3.4.3 Logic Channel Settings”
47
This module cannot be used
because it is not AAF aligned.
Press the
SYSTEM
key to display
the
[Init]
sheet.
Execute
[System Information]
and check the software items.
If
[Not AAF aligned]
is displayed,
send the module for repairs.
—
48
There are module on which AAF
cannot be set to On because of
lack of adjustment.
50
Roll Mode is not available.
Roll Mode cannot be used when
Overlay is enabled.
Recording (Roll Mode)” (p. 147)
51
Cannot use. (Pre Trigger)
The pre-trigger function is not
available when using external
sampling.
“17 External Control” (p. 335)
52
Cannot use. (Roll Mode, Memory
Division)
If the waveform calculation
function is in use, these functions
cannot be used.
53
Cannot use. (Roll Mode, Memory
Division, Wave Calculation)
If one function is in use, the other
functions cannot be used.
Recording (Roll Mode)” (p. 147)
54
Cannot use.(Overlay)
The overlay function is not
available when using the Roll
mode.
Over Past Waveforms (Overlay)”
55
Cannot use. (Overlay, Memory
Division, Wave Calculation)
If roll mode is in use, these
functions cannot be used.
Recording (Roll Mode)” (p. 147)
56
Real-time printing is not available. Recording Length (Shot) is set
to Continuous (
[Cont.] [On]
).
Real-time printing is not available
when using the Recorder function
and a fast timebase setting.
“3.3.3 Recording Length (Number
“6.2 Auto-print Settings” (p. 109)
19
Maintenance and Service
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......