Install or Remove Modules
2.1 Install or Remove Modules
Read “Handling the Instrument and Modules” (p. 7) carefully.
Modules specified when the instrument is ordered are supplied preinstalled. Use the following procedures
to add, replace, or remove modules from the instrument.
•
Up to two logic units can be installed as units 1 to 8. Logic units installed as the other units
cannot be used. Other logic units being installed will be disabled.
•
For information on the analog channel resolution when logic channels are used, refer to “8.10
Detailed Settings for Modules” (p. 164).
To Install a Module
Back side
Required items: One Phillips-head screwdriver (No. 2)
1
Turn off the power to the instrument.
2
Correctly orient the module and insert it firmly all the way in.
Make certain that the labels on the module's panel face the same
direction as the labels on the back side of the instrument.
3
Using a Phillips screwdriver, tighten the two module mounting
screws (M3 × 7.5 mm).
To Remove a Module
Handle
Handle
(Example: 8966)
Back side
Required items: One Phillips-head screwdriver (No. 2)
1
Turn off the power to the instrument.
2
Remove any connection cables or thermocouples connected to
the module.
3
Remove the power cord.
4
Using a Phillips screwdriver, loosen the two module mounting
screws (M3 × 7.5 mm).
5
Grasp the handles and pull the module out.
If not Installing Another Module After Removal
Back side
Blank
panel
1
Install a blank panel.
2
Using a Phillips screwdriver, tighten the two mounting screws
(M3 × 7.5 mm).
Measurements made without a blank panel installed may fail to meet
specifications because of temperature instability within the module.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......