195
Logic Signal Triggers (Logic Trigger)
9.4 Logic Signal Triggers (Logic Trigger)
Follow the steps for below to configure logic triggers. These settings can be configured from the
[Logic
Trig]
sheet of the Trigger settings window.
•
Input signals on logic channels serve as the trigger source.
Triggering occurs when the specified trigger pattern and logical probe combination criteria (AND/OR)
are met.
• The trigger detection method can be selected according to whether a trigger is applied or not when the
criteria are already met at the start of measurement.
•
By using the trigger filter, triggering can be limited so as to occur only when trigger criteria are met for at
least the specified filter width.
Procedure
Press the
DISP
key to open the Waveform screen, and then press the
TRIG.SET
key to open the trigger settings
window (
[Logic Trg.]
sheet).
1
Move the cursor to the channel for which you
want to change the settings.
1.
Logic Channels
2.
3.
2
Use the
F
keys to select the settings.
1. Trigger
Sets the trigger probe logic (AND/OR).
Off
Logic triggering is disabled.
(Default setting)
OR
Triggering occurs when the input signal logic matches any setting in the trigger
pattern.
AND
Triggering occurs only when the input signal logic matches all settings in the
trigger pattern.
2. Filter
Sets the filter width (trigger filter) for triggering. (If necessary)
Suppresses triggering from noise. (p. 193)
Off
Trigger filtering disabled.
(Default setting)
0.1 to 10
Trigger filtering enabled.
The filter width is set as a number of divisions.
Off
Trigger filtering disabled.
(Default setting)
On
Trigger filtering enabled.
Filter width 10 ms
(5 ms when the sampling rate is 100 ns/S)
9
Trigger Settings
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......