Modules (Measurement amplifiers)
These are installed by insertion into the compartments on the right side of the instrument. Modules can
be swapped out as needed.
Application
Model
Number of
channels
Maximum
sampling
rate
A/D
resolution
Maximum
input voltage
Maximum rated
voltage
to earth
Voltage
measurement
Model 8966
Analog Unit
2
20 MS/s
12 bits
400 V DC
300 V AC/DC
(CAT II)
Model 8968
High Resolution Unit 2
1 MS/s
16 bits
400 V DC
300 V AC/DC
(CAT II)
Model MR8990
Digital Voltmeter Unit 2
500 S/s
24 bits
500 V DC
300 V AC/DC
(CAT II)
RMS voltage
measurement
Model 8972
DC/RMS Unit
2
1 MS/s
12 bits
400 V DC
300 V AC/DC
(CAT II)
Model U8974
High Voltage Unit
2
1 MS/s
16 bits
1000 V DC
700 V AC
1000 V AC/DC
(CAT III)
600 V AC/DC
(CAT IV)
Temperature
(Thermocouple)
measurement
Model 8967
Temp Unit
2
-
16 bits
-
300 V AC/DC
(CAT II)
Frequency,
count, pulse duty,
and pulse width
measurements
Model 8970
Freq Unit
2
-
16 bits
400 V DC
300 V AC/DC
(CAT II)
Current
measurement
Model 8971
Current Unit
2
1 MS/s
12 bits
-
Not insulated.
Strain (Strain
gauge type
converter)
measurement
Model 8969
Strain Unit
2
200 kS/s
16 bits
-
33 V rms /
70 V DC
Model U8969
Strain Unit
30 V rms /
60 V DC
Digital signals
and contact signal
measurement
Model 8973
Logic Unit
16
20 MS/s
-
-
Not insulated.
Refer to “18.6 Module Specifications” (p. 360).
Memory
HiCorder
Module
GND
Maximum rated
input voltage
H
L
Maximum rated voltage
to earth
Appx.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......