55619 – PERISTALTIC PUMP SERVICE KIT

To suit Pool Lab ASP (Auto Sampling Photometer) and Pool Lab Hybrid Systems

acid and liquid chlorine feed peristaltic pumps.   Replace every 12 months

PACKAGE CONTENTS

Qty

Part No. / Description

Image

1

DRUM CAP ASSEMBLED
WITH TUBING AND WEIGHTS

CONSISTS OF:
5.0 meters x Part No. 55621
PERI PUMP CLEAR TUBING

1 x Part No. 55620
CAP ASSEMBLY

4 x Part No. 55617-CW
CERAMIC WEIGHT

1 x Part No. 55617-WR
WEIGHT RETAINER (RED)

1

32554-12-BLACK / TUBE & ROLLER KIT

1

55506-BLACK / INJECTOR VALVE

1

55615-LUB / 5g SILICONE LUBRICANT

Rev 6

Summary of Contents for 55619

Page 1: ...12 months PACKAGE CONTENTS Qty Part No Description Image 1 DRUM CAP ASSEMBLED WITH TUBING AND WEIGHTS CONSISTS OF 5 0 meters x Part No 55621 PERI PUMP CLEAR TUBING 1 x Part No 55620 CAP ASSEMBLY 4 x Part No 55617 CW CERAMIC WEIGHT 1 x Part No 55617 WR WEIGHT RETAINER RED 1 32554 12 BLACK TUBE ROLLER KIT 1 55506 BLACK INJECTOR VALVE 1 55615 LUB 5g SILICONE LUBRICANT Rev 6 ...

Page 2: ...MICALS OR TOOLS CLOTHING SKIN EYES ETC EXPOSED TO CHEMICALS THE POOL LAB PERISTALTIC PUMP HAS BEEN DESIGNED FOR USE WITH 16 5 HYDROCHLORIC ACID ONLY USE OF HIGHER STRENGTH HYDROCHLORIC ACID OR OTHER ACID TYPES MAY DAMAGE EQUIPMENT AND RESULT IN A SAFETY HAZARD WARNING Extreme care must be taken when working with the acid feed tubing as under certain conditions the tubing may be pressurized Loosen ...

Page 3: ...ult in a safety hazard the drum must be located within an additional containment vessel with a capacity no less than 200 of the acid drum volume The containment vessel must be made from a plastic suitable for containment of hydrochloric acid eg High Density Polyethylene aka HDPE and should ideally be located directly below the acid pump and feed tubes so that it may catch any potential leaks from ...

Page 4: ...th caution when handling liquid chlorine and or acid while servicing the dosing system 2 To prepare the system for servicing turn OFF and disconnect the filtration pump Relieve the plumbing of any pressure or vacuum a Change the FILTRATION MODE to OFF using the MODE button b Disconnect the filtration pump power cord to prevent accidental start up c If the acid injection point is below water close ...

Page 5: ...d add a small amount of lubricant to the outer face f Lubricate rollers and squeeze lube liberally with silicone lubricant g Install the squeeze tube Ensure tube connectors have the square lugs facing outwards h Replace cover Position top edge in place first then apply downward and inward force i Replace 2 x screws to secure cover When looking at the front of the pump the inlet is on the left and ...

Page 6: ...ad to the pump outlet socket 9 Start the system and check for leaks a Change filtration mode to ON using the MODE button b Ensure water is flowing normally and pressure gauges reading normal c Check plumbing feed tubing and connectors for any signs of leaks 10 Test pump operation with water using the prime function a Place the inlet end of the acid feed tube in a bucket of fresh water b On the chl...