Pre-measurement Inspection
3.2 Pre-measurement Inspection
Verify that the instrument operates normally to ensure that no damage occurred during storage or
shipping. If you find any damage, contact your authorized Hioki distributor or reseller.
(1)
Peripheral Device Inspection
Is the insulation of the probe or connection
cable to be used damaged, or is bare metal
exposed?
Is the clamp cracked or damaged?
Do not use if damage is present, as
you could receive an electric shock.
Request repairs.
When using probes and connection cables
When using a clamp sensor
Metal Exposed
Yes
No
No Metal Exposed
Go to Step (2)
Go to Step (2)
(2) Instrument and Module Inspection
Is there any visible damage to the instrument
or modules?
Do the fans turn and the Hioki logo appear on
the screen?
Does the Waveform screen appear?
If damage is evident, request repairs.
The power cord may be damaged,
or the instrument may be damaged
internally. Request repairs.
The instrument may be damaged
internally. Request repairs.
Yes
No
Nothing appears,
or the display is
abnormal
No
Yes
Yes
When turning power on
Inspection complete
3
Measurement
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......