HI
-
700HS service manual
-
4 –
servicing
4
-
35 Issue 2 06/2020
Re
-
calibration of scale
-
using Console software
(dynamic weighing type)
IMPORTANT: This operation to be undertaken by the distributor or other
qualified persons using traceable calibrated weights.
Before starting the calibration, ensure the scale capacity options are set correctly
To enter the scale menu the ADC security switch needs to be set. See page 4
-
38
Span weight –
enter the weight in grams of the
weight used to calibrate the scale.
Capacity –
The maximum capacity of the scale.
Interval –
The scale interval (in mg).
Re calibrate Zero
This re
-
zero function may be required if the scale has
drifted too far from allowing the operation screen
re
-
zero to work.
In scale settings, go to calibration.
Re
-
calibration of zero is requires no weights.
Press Re calibrate Zero –
ensure there is nothing on
the scale and infeed sensor is clear. Allow machine
to re
-
calibrate zero point.
Wait for message prompt to indicate completion.
Once complete, close and exit scale menu. Save
changes on exiting.
Summary of Contents for HI-700HS
Page 1: ...HI 700HS service manual Issue 2 06 2020 HI 700HS Service manual...
Page 98: ...HI 700HS service manual 2 component parts 2 64 Issue 1 02 2020 HS blow applicator Pneumatics...
Page 238: ...HI 700HS service manual 4 servicing 4 129 Issue 2 06 2020 Side labeller applicator Pneumatics...
Page 270: ...Main electrical cabinet P5 2...
Page 271: ...Main electrical cabinet P5 3...
Page 272: ...Main electrical cabinet display enclosure P5 4...
Page 273: ...Single Servo indexer contol box P5 5...
Page 274: ...Single Servo indexer control box P5 6...
Page 275: ...Triple Servo indexer control box P5 7...
Page 276: ...Triple servo indexer control box P5 8...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ...Low voltage 24v power lift P5 13...
Page 282: ...240v power lift P5 14...
Page 283: ...240v power lift P5 15...
Page 288: ...P5 20 15 touch screen assembly not HS style...
Page 289: ......