HI
-
700HS service manual
-
4 –
servicing
4
-
33 Issue 2 06/2020
Re
-
calibration of scale
-
using Worldview software
(dynamic weighing type)
Calibration
Go to service menu / scale settings
-
must operate security switch in ADC
to allow entry
-
see page 4
-
38
The calibration tab allows calibration and
zeroing of the scale.
To start the calibration process of the
scale, touch Span
Wait for the on
-
screen instruction
IMPORTANT: This operation to be undertaken by the distributor or other
qualified persons using traceable calibrated weights
Clear scale
-
Ensure scale is completely clean with nothing resting on it
Touch OK
Load span weight
-
Place the pre
-
determined calibration weight on the scale
Touch OK
Span complete
When this message appears, the calibration process is
finished, press OK and remove calibration weights
Summary of Contents for HI-700HS
Page 1: ...HI 700HS service manual Issue 2 06 2020 HI 700HS Service manual...
Page 98: ...HI 700HS service manual 2 component parts 2 64 Issue 1 02 2020 HS blow applicator Pneumatics...
Page 238: ...HI 700HS service manual 4 servicing 4 129 Issue 2 06 2020 Side labeller applicator Pneumatics...
Page 270: ...Main electrical cabinet P5 2...
Page 271: ...Main electrical cabinet P5 3...
Page 272: ...Main electrical cabinet display enclosure P5 4...
Page 273: ...Single Servo indexer contol box P5 5...
Page 274: ...Single Servo indexer control box P5 6...
Page 275: ...Triple Servo indexer control box P5 7...
Page 276: ...Triple servo indexer control box P5 8...
Page 277: ......
Page 278: ......
Page 279: ......
Page 280: ......
Page 281: ...Low voltage 24v power lift P5 13...
Page 282: ...240v power lift P5 14...
Page 283: ...240v power lift P5 15...
Page 288: ...P5 20 15 touch screen assembly not HS style...
Page 289: ......