4.
Close door and latch, as follows:
a. Close door and hold in a closed position
by grasping door and instrument case
with one hand.
b. Gently lower latch.
• Safety clamp device is automatically
disengaged.
2-4
Getting Started
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
c. Press safety clamp fitment into recess below
mechanism.
d. Using a finger tip, firmly push tubing toward back of
Air-in-Line (
AIL
) Detector.
Safety Clamp Fitment
Push tubing toward
back of AIL Detector.
Loading
(Continued)
•
Do not touch
the administration
set while closing the door.
Failure to follow this instruction
may result in infusion rate
inaccuracy.
• To
prevent a potential free-flow
condition
, ensure no extraneous
object (for example, bedding,
tubing, glove) is enclosed or
caught in the Pump Module door.
WARNINGS
Preparing Administration Set (
Pump Module
)
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
To reduce the potential for nuisance
AIL
alarms,
ensure tubing is fully
inserted
in
AIL
Detector.
CAUTION
5.
Open roller clamp.
6.
Verify no fluid is flowing through drip chamber.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......