Auto-Restart
Part of Alaris
®
System’s Downstream Occlusion Detection system
designed to minimize nuisance, patient-side occlusion alarms.
Allows system to automatically continue an infusion following
detection of a patient-side occlusion if downstream pressure falls to
an acceptable level within a
15
-second "Checking Line" period. If
this feature is enabled, "Checking Line" function occurs when
downstream pressure exceeds pressure limit.
• In Selectable Pressure Mode: Pressure limit will be either user
adjustable or "locked" in system configuration.
• In Pump Pressure Mode: Pressure limit is a function of flow rate
and is automatically determined by device.
If downstream pressure decreases to a predetermined level, (below
50%
of pressure limit) during
15
-second "Checking Line" period,
infusion automatically continues. If condition is not cleared within
15
seconds, a "Partial Occlusion - Patient Side" alarm occurs.
Using Guardrails
®
Editor Software, system can be configured to allow
0
(zero) to
9
restart attempts within a rolling
10
minute period. If
allowable number of restarts is exceeded or if feature is set to zero,
an "Occluded - Patient Side" alarm occurs when system detects
downstream pressure that exceeds pressure limit.
Default Occlusion Pressure
Starting occlusion pressure limit which can be configured by profile in
25
mmHg increments.
Free Flow Protection
All Alaris
®
Pump Module/Gemini Infusion System administration sets
utilize a unique clamping device (safety clamp on pumping
mechanism) to prevent inadvertent free-flow when administration set
is removed from instrument.
KVO Rate Adjust
Used to select
KVO
(Keep Vein Open) rate (
0.1
to
20
mL/h allowed),
which is rate of fluid flow after an "Infusion Complete" occurs.
KVO
rate will never exceed infusion rate.
Pump Module
Features and Displays
(Continued)
Features and Definitions
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
General Information
2-83
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......