2-50
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
1.
Set up infusion as described in "Continuous Infusion -
Drug Calculation" procedure, but do not start infusion.
2.
Press
BOLUS
soft key.
3.
To enter bolus dose, use numeric data entry keys.
• After a bolus dose and weight (if used) are entered,
bolus
VTBI
and concentration [conc] alternate in Main
Display.
Bolus Dose
4.
Press soft key next to appropriate unit of measure for
dose.
5.
To enter bolus duration, use numeric data entry keys.
•
TOTAL DOSE
alternates with
INFUSE AT
rate.
6.
Verify parameters are correct and press
START
soft key.
A
A
A
DOSE
PATIENT
WEIGHT
BOLUS DOSE
Drug Calculation
_ 2000
mcg
mcg/kg
mg
DURATION
mg/kg
[Conc]: 5000 mcg/mL
SETUP
>Select the Desired Dosing
Units
CONT-
INUOUS
NOTES:
If
mcg
or
mg
is selected as the dosing unit, a
PATIENT
WEIGHT
entry cannot be made. If
mcg/kg
or
mg/kg
is
selected as the dosing unit, a
PATIENT WEIGHT
entry is
required.
To see details during the bolus infusion, press the
CHANNEL
SELECT
key.
Stopping and Restoring Bolus Dose
Reference "Primary Infusion - With Guardrails
Protection", "Bolus Dose" procedure.
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Infusion - NO
Guardrails
®
Suite MX Protection
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......