2-98
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Specifications
(Continued)
Critical Volume:
Maximum over-infusion which can occur in the event of a single-fault
condition will not exceed
2%
of nominal syringe fill volume during loading
and
1%
of maximum syringe travel after syringe loading.
Dimensions:
4.5"W
x
15.0"H
x
7.5"D
Environmental Conditions:
Operating
Storage/Transport
Temperature Range:
41
-
104°F
-
4
-
140°F
(
5
-
40°C
)
(-
20
-
60°C
)
Relative Humidity:
20
-
90%
5
-
85%
(Avoid prolonged exposure
Noncondensing
Noncondensing
to relative humidity >85%)
Atmospheric Pressure:
525
-
4560
mmHg
375
-
760
mmHg
(
700
-
6080
hPa)
(
500
-
1013
hPa)
Equipment Orientation:
To ensure proper operation, PC Unit must remain in an upright position.
Flow Rate Programming:
Flow rate range is from
0.01
to
999
mL/h and can be selected as follows:
Flow Rates (mL)
Selectable Increments (mL/h)
0.01
-
9.99
0.01
10
-
99.9
0.1
100
-
999
1
Rate Restriction by Syringe Size:
Syringe Size (mL)
Flow Rate Range (mL/h)
50
/
60
0.1
-
999
30
0.1
-
650
20
0.1
-
500
10
0.1
-
250
5
0.1
-
150
3
0.01
-
100
1
0.01
-
30
Fluid Ingress Protection:
IPX1
, Drip Proof
Syringe Module
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......