General Information
1-31
Alaris
®
System Directions for Use
PC Unit Section
Battery Operation
:
Battery run time is a function of the number of modules attached and module
activity. With a new, fully charged battery, the system will operate as follows
before a "
BATTERY DISCHARGED
" message occurs:
•
8
hours with
1
Pump Module infusing at
25
mL/h
•
6
hours with
1
Pump Module infusing at
25
mL/h and
1
Auto-
ID
Module
•
4
hours with
4
Pump Modules infusing at
25
mL/h
•
3
hours with
4
Pump Modules infusing at
25
mL/h and
1
Auto-
ID
Module
•
6
hours with
1
active
SpO
2
Module
•
8
hours with
1
Syringe Module or
PCA
Module infusing at
5
mL/h
•
4
hours with
4
Syringe Modules infusing at
5
mL/h
•
5.5
hours with
1
active
E
t
CO
2
Module
When using the Communications Interface Board Accessory, battery run time
may be reduced by
25%
.
Communication Data Port
:
RS-232
with an
RJ45
connector.
Dimensions
:
6.9"W
x
8.8"H
x
9"D
(including pole clamp)
Electric Classification
:
Class
1
, Internally Powered Equipment
Electronic Memory
:
System configuration parameters stored in volatile memory are retained for at
least
6
months by internal backup lithium battery. Module-specific parameters are
stored for
8
hours when system is turned off. After
8
hours of continuous off-time,
or if a module is detached, module-specific trend data (if applicable) and module-
specific operating parameters are automatically purged. If an Alaris
®
PCA, SpO
2
or
E
t
CO
2
Module is detached and replaced with another
PCA, SpO
2
or
E
t
CO
2
Module,
its module-specific trend data is purged.
Environmental Conditions
:
Operating
Storage/Transport
Temperature Range:
41
-
104°F
-
4
-
140°F
(
5
-
40°C
)
(-
20
-
60°C
)
Relative Humidity:
20
-
90%
5
-
85%
(Avoid prolonged exposure
Noncondensing
Noncondensing
to relative humidity >85%)
Atmospheric Pressure:
525
-
4560
mmHg
375
-
760
mmHg
(
700
-
6080
hPa)
(
500
-
1013
hPa)
Equipment Orientation
:
To ensure proper operation, the instrument must remain in an upright position.
Fluid Ingress Protection
:
IPX1
, Drip Proof
Leakage Current
:
Less than
100
microamps
Power Requirements
:
100
-
240V
~,
50/60
Hz,
150 VA MAX
Weight
:
7.2
lbs
Specifications and Symbols
Specifications
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......