Troubleshooting and Maintenance
2-117
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Messages
(Continued)
Alarms, Errors, Messages
(Continued)
Checking Line
Secondary
Patient-side occlusion occurred;
Auto-Restart feature monitoring
downstream pressure to determine
if infusion can continue.
Secondary infusion in progress on
indicated module.
None
None. When secondary
VTBI=
"
0
",
infusion will revert to programmed
primary parameters.
Message
Meaning
Response
Pump Module
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
After Call Back
NEOI (Near End of Infusion)
Syringe Not Recognized
Infusion completed.
Syringe almost empty.
Installed syringe of unknown type
and size.
Press
CONFIRM
soft key.
None. This is a timed event that
can be set. To set or change this
option, reference "General
Information", "Configurable
Settings".
Select and confirm correct syringe
type and size, and then press
CONFIRM
; or use a syringe type
and size that system can
automatically and correctly identify.
Message
Meaning
Response
Syringe Module
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......