Regulations and Standards
A-9
Alaris
®
System Directions for Use
Appendix Section
IEC 61000-4-6
Conducted RF
IEC 61000-4-3
Radiated RF
Portable and mobile RF communications equipment
should be used no closer to Alaris
®
System
(including cables) than recommended separation
distance calculated from equation applicable to
frequency of transmitter.
Recommended Separation Distance
12
— d = [-----]
√
P
V
2
12
— d = [-----]
√
P 80 MHz - 800 MHz
E
1
12
— d = [-----]
√
P 80 MHz - 2.5 GHz
E
1
d = recommended separation distance in
meters (m).
P = maximum output power rating of transmitter in
watts (W) according to transmitter manufacturer.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than compliance level in each
frequency range.
Interference may occur in vicinity of equipment
marked with following symbol:
3 Vrms
150 kHz - 80 MHz
3 V/m
80 MHz - 2.5 GHz
10 Vrms
10 V/m
Table 3
Electromagnetic Immunity - Life Support Equipment
Immunity Test
Electromagnetic Environment - Guidance
IEC 60601-1-2
Test Level
Compliance
Compliance
(Continued)
Alaris
®
System (Continued)
Electromagnetic Environment
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......