Battery Discharged
Channel Disconnected
Very Low Battery <5 minutes to
system shutdown
Operation of all modules stopped
due to insufficient battery charge.
Module(s) disconnected while in
operation or have a communication
problem.
Battery has
5
minutes or less of
power at current power
consumption rate before operation
stops.
Connect AC power cord to power
source; alarm will be silenced.
Press
RESTART
key on affected
module to continue operation of
paused modules.
To silence alarm and clear
message from screen, press
CONFIRM
soft key. Reattach
module, if desired, ensuring it is
securely "clicked" into place at
Module Release Latch. If alarm is
still present, replace module with
an operational instrument.
Connect
AC
power cord to power
source; alarm will be silenced.
Alarms
Errors
Alarms, Errors, Messages
(Continued)
Alarm
Meaning
Response
Audio System Error
Channel Error
Main speaker failure.
Error detected. Operation stops on
affected module.
1. Visually check alarm status to
determine whether or not an
operational alarm also needs to
be addressed (red Alarm Status
Indicator would be lit).
2. Press
SYSTEM ON
key to power
down system. Replace PC Unit
with an operational instrument.
Service by qualified personnel
is required.
To silence alarm and continue
operation of unaffected modules,
press
CONFIRM
soft key . Replace
module with an operational
instrument, as required. Service by
qualified personnel is required.
Error
Meaning
Response
Troubleshooting and Maintenance
1-37
Alaris
®
System Directions for Use
PC Unit Section
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......