Troubleshooting and Maintenance
1-39
Alaris
®
System Directions for Use
PC Unit Section
Alarms, Errors, Messages
(Continued)
Plug the PC Unit into an
AC
outlet during storage, to ensure a
fully charged battery when needed.
AC
indicator light (
) will
be on whenever the PC Unit is plugged in.
Storage
Messages
Battery Run Time = X.X hours
Low Battery
Panel Locked
Panel Unlocked
Powering Down
Replace Battery
AC
power cord is disconnected
from power source. Approximate
remaining battery run time under
current power consumption rate is
displayed.
Low battery threshold sensed;
remaining battery run time is
limited.
Tamper Resist feature is active and
key was pressed.
Tamper Resist feature deactivated.
Last module powering off. System
shuts off in indicated number of
seconds.
Occurs at System On. Battery has
less than
50%
of original capacity.
None. Connect
AC
power cord to
power source as soon as possible.
Connect
AC
power cord to power
source; alarm will be silenced.
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
Press any key, except
SYSTEM ON
key, to cancel power down
sequence.
Press either
SYSTEM OFF
or
CONFIRM
soft key to continue
normal operation with reduced
battery capacity. Service by
qualified personnel is required.
Message
Meaning
Response
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......