5.
Verify parameters are correct and press
NEXT
soft key to
confirm.
• If programmed total dose drug amount is outside Soft
Limit for that care area, a prompt appears before
programming can continue. If
Yes
soft key is pressed,
programming continues; if
No
soft key is pressed,
infusion needs to be reprogrammed.
•
If programmed total dose drug amount is outside Hard
Limit for that care area, a prompt appears before
programming can continue. Infusion needs to be
reprogrammed.
•
If a dose outside of Soft Limits has been entered and
verified as correct, Message Display also shows either
"
LLL
" for a low dose or "
↑↑↑
" for a high dose.
6.
To enter secondary infusion
VTBI
, press
VTBI
soft key and
use numeric data entry keys.
2-42
Programming
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Infusion
(Continued)
• If selected drug is not weight-based,
Not Used
displays in
PATIENT WEIGHT
field.
• If hospital/facility practice guidelines identify selected
drug as weight-based, prompt for a patient weight in
kilograms or
BSA
appears (as in illustrated example,
which reflects use of Methotrexate).
[Conc]: 342 mg/mL
_ _ _
mL/h
_ _ _ _
h
min
RATE =
VTBI
DURATION
RATE
VOLUME
A
>Enter Secondary VTBI
DOSE =
SETUP
1200
mg/m
2
ALL
Guardrails Drug Setup
Methotrexate
SECONDARY
Secondary Infusion
- With Guardrails
®
Suite MX Protection
(
Pump Module
)
(Continued)
A
DRUG
AMOUNT
DILUENT
VOLUME
6840
mg
20
mL
[Conc]: 1.7 mg/mL
>Enter BSA
DOSE =
_ _ _
mg/m
2
Guardrails Drug Setup
Methotrexate
BSA
_ _ _
m
2
DRUG
LIBRARY
PRIMARY
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......