Programming
2-67
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Channel Options
Guardrails Drugs
Pressure Limit - P
Channel Labels
A
>Select an Option or EXIT
EXIT
Multidose
Guardrails IV Fluids
>Select an Option or EXIT
EXIT
Pressure Limit Selection
Selectable
Pump
A
3.
Press
Pressure Limit
soft key.
4.
Press either
Pump
or
Selectable
pressure soft key. If
Selectable
is pressed, continue with next step; otherwise,
Selecting Pressure Limit
(Continued)
Pump Module
(Continued)
7.
Press
START
soft key.
Pressure Limit Selection
Selectable Pressure
525
mmHg
Up
Down
CONFIRM
A
5.
To select occlusion pressure limit, press either
Up
or
Down
soft key.
6.
Verify correct occlusion pressure limit input and press
CONFIRM
soft key.