Getting Started
2-5
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
3.
Remove set, as follows:
a. Gently pull tubing below Air-in-Line Detector forward
and out.
b. Lift upper fitment from upper fitment receptacle.
4.
If set is being removed to begin a gravity flow:
a. Depress blue ridged release tab on upper side of
safety clamp device.
b. Slide white slide clamp into blue fitment (open
position).
c. Adjust flow rate using set’s roller clamp.
Removing
1.
Close roller clamp.
2.
Open Pump Module door.
• Set’s safety clamp fitment automatically closes to
prevent accidental free-flow.
White Slide Clamp
(shown in closed position)
1.
Prepare primary solution container in accordance with
manufacturer’s directions for use.
2.
Open administration set package, remove set, and close
roller clamp. (Reference set’s directions for use.)
3.
Insert administration set spike into prepared fluid
container, following accepted hospital/facility procedure,
and hang container
20
inches above Pump Module.
4.
Fill drip chamber to
2
/
3
full.
5.
If container requires venting, open vent cap on
administration set spike.
Priming
Preparing Administration Set (
Pump Module
)
(Continued)
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......