The Dose Request Cord can be configured to provide both
audio and visual prompts to the patient. Visual prompts are
provided through the
LED
indicator on the Dose Request
Cord. Default configuration for the Dose Request Cord is
established in the system configuration.
To change Dose Request Cord configuration:
1.
Press
CHANNEL SELECT
key.
2.
From Main Display, press
OPTIONS
key.
3.
Press
Dose Request Setup
soft key.
3-26
Programming
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Configuring Dose Request Cord
A
>Select an Option or
EXIT
EXIT
Dose Request Setup
Infusion Modes
Patient Histor
y
PCA Pause Limits
Drug Event
Histor
y
Channel Options
1 of 2
PAGE
DOWN
Infusion Modes
(Continued)
A
TIME
DRUG EVENT
2003-06-10
07:08
07:08
06:10
09:01
PCA Dose, 1 mg
Continuous, 2 mg/h
Bolus Dose stopped
08:38
PCA + Cont infusion
08:08
New s
y
ringe
Unmet demands - 500
Bolus Dose, 1mg
Histor
y
cleared
06:08
06:00
Drug Event Histor
y
1
EXIT
P
AGE
DOWN
>Press PAGE DOWN or
PAGE UP to View Events
Morphine
NOTE:
The
Drug Event History
stores approximately
12
hours of
events and is automatically cleared upon selection of
New
Patient?
,
Yes
during start-up or upon changing drug in Drug
Library.
Viewing Drug Event History
(Continued)
4.
To scroll through history, press
PAGE DOWN
soft key.
5.
To return to Main Display, press
EXIT
soft key.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......