Getting Started
2-15
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
1.
Press
OPTIONS
key.
2.
Press
Prime Set with Syringe
soft key.
Channel Options 1 of 2
A
>Select an Option or EXIT
EXIT
PAGE
DOWN
Guardrails Drugs
Pressure Limit - High
Multidose
Prime Set with S
y
ringe
Guardrails IV Fluids
3.
Press and hold
PRIME
soft key until fluid flows and
priming of syringe administration set is complete.
4.
When priming is complete, release
PRIME
soft key.
5.
To return to main screen, press
EXIT
soft key.
NOTES:
When manually priming (per hospital/facility protocol) and an
administration set having a pressure sensing disc is in use,
depress the disc between
2
fingers while priming and prime
uphill (distal end of pressure sensing disc/tubing pointing
upward).
Fluid is delivered during priming only while the
PRIME
soft key
is pressed. Each press of the
PRIME
soft key delivers up to
2
mL of priming fluid per continuous press. To deliver
additional amounts, press the
PRIME
soft key again.
Volume used during priming is displayed but not added to
VTBI
.
Pressure Sensing Disc
Not
Installed
Priming
(continued)
Preparing Syringe and Administration Set (
Syringe Module
)
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......