Troubleshooting and Maintenance
2-113
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Alarms, Errors, Messages
(Continued)
Alarms
(Continued)
Syringe Empty
Syringe is empty.
¤
If syringe is not empty, other
possibilities are:
• Pressure sensing disc
¤
inappropriate/defective.
• Syringe plunger travel
¤
impeded.
• Pressure transducer defective.
¤
Set up new infusion or press
CHANNEL OFF
key.
Verify appropriate pressure sensing
disc is in use and functioning
properly.
Verify syringe plunger movement is
unimpeded.
If syringe is not empty and above
actions do not correct alarm,
contact qualified service
personnel.
Alarm
Meaning
Response
Syringe Module
(Continued)
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
When a syringe installation problem is detected, a visual
signal is displayed. Text in the display blinks to indicate the
location of the problem.
A
A
A
CONTINUOUS INFUSION
Drug Calculation
Heparin
A
>Press
or Adjust
Syringe
CANCEL
CONFIRM
S
y
ringe Installation
plunger
flange
clamp
lever
Syringe Adjustment Alarms
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
-- Continued on Next Page --
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......