The Pump and Syringe Modules use the following parameters,
entered during the drug calculation setup procedure:
•
Bolus dose duration:
Time period over which bolus dose
is to be administered.
•
Bolus dose units:
Units used in calculating bolus dose.
Bolus dose units are selected from alternatives provided.
•
Diluent volume:
Volume of fluid used as diluent for drug
(mL).
•
Dosing units:
Units used to calculate continuous infusion
drug dose. Dosing Units are selected from alternatives
provided.
•
Drug amount:
Amount of drug in
IV
container (gram, mg,
mcg, mEq, or units).
•
Patient weight:
Weight of patient (kg); this is an optional
parameter that is not needed unless drug dose is
normalized for patient weight.
•
Time units:
Time base for all calculations (minute, hour, or
day).
The bolus dose, drug dose, and flow rate parameters are
calculated using the above parameters, as follows:
• Bolus dose = bolus dose x patient weight (if used).
• Bolus dose administration rate (
INFUSE AT:
):
When duration is entered = total dose / duration in minutes.
When Max Rate is used = Max Rate /
60
x concentration.
• Bolus dose duration = bolus
VTBI
/ bolus rate.
• Bolus dose
VTBI
= bolus dose / drug concentration.
• Bolus rate = bolus
VTBI
/ duration.
• Continuous drug dose = flow rate x drug concentration
(normalized for patient weight if specified by entering a
patient weight).
• Continuous flow rate = drug dose / drug concentration
(normalized for patient weight if specified by entering a
patient weight).
• Duration =
VTBI
/ rate.
• Drug concentration = drug amount / diluent volume.
• Rate =
VTBI
/ duration.
General Information
2-91
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Drug Calculation Definitions and Formulas
The Drug Calculation feature is to
be used only by
personnel
properly trained
in the
administration of continuously
infused medications. Extreme
caution should be exercised to
ensure the correct entry of the drug
calculation infusion parameters.
WARNING
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......