Alarms, Errors, Messages
(Continued)
Troubleshooting and Maintenance
2-111
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Accumulated Air-in-Line
Air-in-Line
Check IV Set
Close Door
Flo-Stop Open - Close Door
Occluded - Fluid Side/Empty
Container
Occluded - Patient Side
Partial Occlusion - Patient Side
A large number of air bubbles
smaller than current air-in-line limit
has recently passed detector.
Air has been detected in
administration set during an
infusion. Infusion stops on affected
module.
Administration set is not properly
installed. Infusion stops on
affected module.
Door opened during an infusion.
Infusion stops on affected module.
Safety clamp device is in open
position while door is open.
Indicates either upstream occlusion
or empty container. Infusion stops
on affected module.
Increased back pressure sensed
while infusing in pump delivery
mode. Infusion stops on affected
module.
Partial occlusion of patient side of
IV line detected by Auto-Restart
feature.
Clear air from line. To continue
infusion, press
RESET
soft key and
then
RESTART
key.
Ensure tubing is properly installed
in Air-in-Line Detector. If air is
present, clear air from
administration set. Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Close roller clamp, remove and
reinstall administration set, close
door, open roller clamp, and then
press
RESTART
key.
Close door. Press
RESTART
key,
or press
CHANNEL SELECT
key
and then
START
soft key.
Close roller clamp on
administration set or close door.
Clear occlusion on fluid side of
instrument. If necessary, refill drip
chamber. Press
RESTART
key
,
or
press
CHANNEL SELECT
key and
then
START
soft key.
Clear occlusion. Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Clear occlusion. Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Alarm
Meaning
Response
Pump Module
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
Alarms
(Continued)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......