2-94
General Information
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Configurable Settings
(Continued)
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
Enabled - Disabled
0.1
-
2.5
mL/h
0.5
-
5%
0.1
-
99.9
mL/h in
0.1
mL/h increments;
100
-
999
mL/h in 1 mL/h
increments
Enabled - Disabled
1
-
60
minutes or
25%
of
remaining infusion time,
whichever comes later
25
-
1000
mmHg in
1
mmHg increments
Low, Medium, High
Enabled - Disabled
Feature
ALL Mode
Auto Pressure
Back Off (
after occlusion
)
Fast Start
KVO (
Keep Vein Open
)
• Rate Adjust
• Volume Adjust
Max Rate
Near End (
NEOI
)
• Alert Time
Occlusion Pressure Set Point:
• With Disc
• No Disc
Priming
Default Setting
Options
Disabled
Disabled
Enabled
Enabled
Disabled
1
mL/h
5%
999
mL/h
Disabled
60
1000
mmHg
High
Disabled
Syringe Module
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......