3-46
General Information
Alaris
®
System Directions for Use
Patient Controlled Analgesia Module Section
Features and Displays
(Continued)
Features and Definitions
(Continued)
Patient History
The PCA Module records and displays patient history for up to
24
hours, and may be trended to following intervals:
1
hr,
2
hr,
4
hr,
8
hr,
12
hr,
24
hr. Patient history includes the following trending
information:
• total demands
• delivered demands
• total drug delivered
• time and date patient history last cleared
• average drug per hour
• total amount of drug delivered via:
¤
PCA
dose
¤
continuous infusion
¤
loading dose
¤
bolus dose
PCA Dose
Enables a patient to self-administer a bolus infusion to be
delivered at programmed lockout intervals through Dose Request
Cord. When programmed in
PCA
+continuous mode, continuous
infusion resumes following
PCA
dose.
PCA Pause Protocol
An optional and hospital-configurable feature intended to align
with hospital/health system’s current protocol for patient
monitoring during
PCA
therapy. When enabled,
PCA
infusion
pauses and alarms when defined monitoring values for Alaris
®
S
p
O
2
and/or Alaris
®
E
t
CO
2
Modules are reached.
Pressure Limit
Downstream occlusion alarm threshold can be set to low, medium
or high. Syringe variability may impact occlusion pressure
sensing. Variability may reduce device’s time to alarm and/or may
require that a higher alarm pressure limit be programmed.
Priming
Allows a limited volume of fluid to be delivered in order to prime
administration set prior to being connected to a patient or after
changing a syringe. When priming, a single continuous press of
PRIME
soft key delivers up to
2
mL of priming/fluid.
Restore
To simplify programming, can be used to recall previous
PCA
programming parameters for same patient. This option is only
available if patient is not new and system is powered up within
8
hours of last usage.
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......