The following procedures should be used only when the drug
to be infused is not listed in the Drug Library. When
programming a drug not listed in the Drug Library, the drug
calculation must be programmed using the
DRUG CALC
soft
key within the Drug Library. There are no limits associated
with any non-library drug calculation.
1.
Perform following steps (reference PC Unit Section of this
, "General Setup and Operation", "Start-Up"):
a. Power on system.
b. Choose
Yes
or
No
to
New Patient?
c. Confirm current profile or select a new profile.
d. Enter patient identifier, if required.
2.
Prepare and load syringe/administration set (reference
"Getting Started").
3.
Prime (reference "Getting Started").
4.
Start applicable infusion, as described in following
procedures:
Basic Infusion
Continuous Infusion - Drug Calculation
Bolus Dose
Programming
2-45
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Infusion - NO
Guardrails
®
Suite MX Protection
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
The following procedures should be used only to set up a
Basic Infusion
. To program an infusion using
Guardrails
Drugs
, reference "Primary Infusion - With Guardrails
The illustrations in this procedure assume the following:
•
ALL
Mode (Syringe Module), Drug Calculation, Dynamic
Pressure Display, Profiles, and Volume Duration
configurable settings are enabled.
•
NEOI (
Syringe Module) and Delay Options configurable
settings are disabled.
1.
Press
CHANNEL SELECT
key.
Basic Infusion
When the
pressure sensing disc
is not being used
and an occlusion
occurs, there is a risk of infusing
pressurized buildup of infusates
upon correction of the occlusion.
To avoid an inadvertent bolus,
relieve the pressure before
restarting the infusion.
WARNING
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......