4-22
General Information
Alaris
®
System Directions for Use
SpO
2
Modules Section
Features and Displays
(Continued)
Features and Definitions
(Continued)
8220 Series
Fast SAT
When Fast
SAT
is enabled and there is
1
data point that is
significantly different from a previous data point, averaging is
disregarded and most recent data point is displayed. For example, if
readings were
97%
,
96%
,
95%
and
85%
, displayed saturation level
would be
85%
.
PI
Perfusion Index (
PI
) is a scaled numeric value derived from
magnitude of pulsations displayed on plethysmographic (pleth)
waveform. It is calculated as a percentage of pulsatile signal to
nonpulsatile signal.
PI
is used to find best perfused site for sensor
placement (larger the
PI
, stronger the perfusion). Operating range is
0.02
to
20
. Desired number is greater than
1
or as large as possible.
Saturation Averaging Time
Averaging time can be set to
2
,
4
,
8
,
10
,
12
,
14
or
16
seconds.
Sensitivity Mode
Sensitivity mode, normal or maximum, of current monitoring
configuration is displayed in options mode. Normal setting is used for
normal patient monitoring purposes. Maximum setting is used for
improved low perfusion performance.
SET
®
Technology
Signal Extraction Technology
®
(
SET
®
) uses adaptive filters to separate
arterial signal from nonarterial noise.
SET
®
provides for accurate
readings under extreme conditions; such as, low perfusion and
motion.
Signal I.Q.
™
Feature
A visual indication of pulsation at sensor site. Vertical bar height
indicates quality of measured signal. Signal
I
.
Q
.
™
feature is related to
proper sensor application, adequate arterial signal and intensity of
motion. Use Signal
I
.
Q
.
™
feature to verify optimal sensor placement.
(8220)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......