Anesthesia Mode
Bolus Dose Complete
Delay Complete
Infusion Complete
Infusion Complete - KVO
Panel Locked
Panel Unlocked
Pause
Start time for next dose has
passed.
Anesthesia Mode discontinued
when disconnected from
AC
.
Module running in continuous
infusion mode if programmed.
Delay time completed.
Current infusion completed.
Programmed volume-to-be-infused
delivered; module running at
KVO
rate.
Tamper Resist feature is active and
a key was pressed.
Tamper Resist feature deactivated.
Pause control pressed; infusion
stopped.
Start of next dose passed.
Press
CONFIRM
soft key.
None
Press
RESTART
key, or press
CHANNEL SELECT
key and then
START
soft key.
Set up a new infusion or press
CHANNEL OFF
key.
Set up a new infusion or press
CHANNEL OFF
key.
If appropriate, deactivate Tamper
Resist feature using Tamper Resist
Control on back of PC Unit.
None.
To resume infusion, press
RESTART
key, or press
CHANNEL
SELECT
key and then
START
soft
key.
Press
CONFIRM
soft key.
2-116
Troubleshooting and Maintenance
Alaris
®
System Directions for Use
Pump and Syringe Modules Section
Alarms, Errors, Messages
(Continued)
Messages
Message
Meaning
Response
Pump and Syringe Modules
CHANNEL
SELECT
CHANNEL
OFF
RESTART
PAUSE
ALARM
INFUSE
STANDBY
ALARM
INFUSE
STANDBY
CHANNEL
SELECT
CHANNEL
OFF
PAUSE
RESTART
RATE (mL/h)
Summary of Contents for Alaris 8000 Series
Page 8: ......
Page 52: ......
Page 54: ......
Page 176: ......
Page 242: ......
Page 284: ......
Page 317: ...6 Alaris System DFU Section SCAN CANCEL READY Alaris Auto ID Module 8600 Series...
Page 318: ......
Page 338: ......
Page 339: ...Max Load 30V 1A NURSE CALL Nurse Call Accessory Model 8010 7 Alaris System DFU Section...
Page 340: ......
Page 350: ......
Page 351: ...Communications Interface CI Board Accessory Model 8012 8 Alaris System DFU Section...
Page 352: ......
Page 361: ...A Alaris System DFU Section Appendix Maintenance Regulations and Standards...
Page 362: ......
Page 377: ......
Page 378: ......